Scope Calibration of Tensile Tester & Universal Testing Machine
1. Force |
|
Parameter / Range / Item |
Calibration measurement capabilities |
(Force, Tension mode) |
|
4 to <100 Newton |
+/- 0.060 Newton |
100 to <1,000 Newton |
+/- 0.47 Newton |
Standard / Technique / Method /Equipment / Remark |
|
ISO 7500-1:2004 (E) /Standard weight. & Standard Loadcell. (Accreditation) |
2. Speed verification |
|
Parameter / Range / Item |
Calibration measurement capabilities |
50 mm./min |
+/- 0.5 mm./min |
100 mm./min |
+/- 0.5 mm./min |
200 mm./min |
+/- 0.5 mm./min |
300 mm./min |
+/- 0.5 mm./min |
400 mm./min |
+/- 0.5 mm./min |
500 mm./min |
+/- 0.5 mm./min |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-17 Verification by Standard stopwatch and Vernier caliper. |
3. Distance (Extensometer) |
|
Parameter / Range / Item |
Calibration measurement capabilities |
100 % (Gauge 2 cm.) |
+/- 1.5 % |
200 %(Gauge 2 cm.) |
+/- 1.5 % |
300 %(Gauge 2 cm.) |
+/- 1.5 % |
400 %(Gauge 2 cm.) |
+/- 1.5 % |
500 %(Gauge 2 cm.) |
+/- 1.5 % |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-17 by direct comparison to standard gauge block. |
* expressed as an uncertainty (+/-) which for k=2, providing a level of confidence of approximately 95%