Scope Calibration of Rheometer (MDR&RPA)
1.Temperature |
|
Parameter / Range / Item |
Calibration measurement capabilities |
(Temperature Accuracy.) |
|
100 to 210 °C |
+/- 0.50 °C |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-07 by direct comparison to standard Thermometer. (Accreditation) |
2. Torque verification |
|
Parameter / Range / Item |
Calibration measurement capabilities |
Degree of Verify |
|
0.5 Dregree. |
+/- 1.60% |
1.0 Dregree |
+/- 1.60% |
3.0 Dregree |
+/- 1.60% |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-15 Verification by Standard Torque |
3. Speed |
|
Parameter / Range / Item |
Calibration measurement capabilities |
(Speed of Die) |
|
100 cpm |
+/- 0.1 cpm |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-17 by measurement with Standard Stopwatch. |
* expressed as an uncertainty (+/-) which for k=2, providing a level of confidence of approximately 95%
Scope Calibration of Rheometer (Only ODR type)
1.Temperature |
|
Parameter / Range / Item |
Calibration measurement capabilities |
(Temperature Accuracy.) |
|
100 to 200 °C |
+/- 0.40 °C |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-06 by direct comparison to standard Thermometer. (Accreditation) |
2. Torque verification |
|
Parameter / Range / Item |
Calibration measurement capabilities |
Degree of Verify |
|
0.5 Dregree. |
+/- 1.60% |
1.0 Dregree |
+/- 1.60% |
3.0 Dregree |
+/- 1.60% |
Standard / Technique / Method /Equipment / Remark |
|
In-house method WI 7.2-15 Verification by Standard Torque |
* expressed as an uncertainty (+/-) which for k=2, providing a level of confidence of approximately 95%